The single valve turns a total of nine and a half times; the first two and a half turns utilize the slight, minute changes for the pediatric animal patient while the remaining rotations are to be used with the adult and extra-large sized bellows and housings.
Low 50 psi Supply Gas Alarm – The Low Supply Gas Alarm is activated when the supply gas pressure drops below 35 psi (2.4 bar). During an alarm condition, the red light on the front panel is illuminated and a steady, continuous tone is heard. This alarm automatically resets when the pressure increases above 40 psi (2.7 bar).
All Hallowell Ventilators are equipped with dual airway pressure alarms that monitor and sound when the Peak Inspiratory Pressure (PIP) of any breath either exceeds the preset Maximum Working Pressure Limit (MWPL) or fails to reach 6 cm H2O as in the case of a disconnect.
Volume – Cycled microprocessor-controlled ventilator.
Time cycled – Ensures consistently spaced breaths. The possibility of stacking breaths or auto-PEEP is eliminated by keeping the I:E ratio constant at 1:2. This means that regardless of the respiratory rate, setting two-thirds of the breath will remain for expiration. The respiratory rate is adjustable from 6 – 40 breaths per minute (BPM).
Volume constant – Ensures that each delivered breath will have essentially the same volume, independent of changes in patient compliance and airway resistance. The deliverable tidal volume (TV) range is from approximately 20 cc to 3,000 cc.
Pressure Limited – Provides for patient safety by limiting the maximum working pressure (MWPL) to a user adjustable pressure range of 10 – 60 cm
What makes the 2024IE PRO different from the Standard 2024 PRO?
Adjustable I:E Ratio – Normally, without this option, the I:E Ratio is fixed at 1:2, that is, 1/3 of the period of each breath is inspiration and 2/3 of each breath is exhalation. Regardless of the respiratory rate, the animal patient will always have 2/3 of the cycle to exhale. This I:E ratio can also be adjusted to a fixed value other than 1:2 at no extra charge.
The adjustable range is from 1:1 to 1:4. The pros and cons are that this option affords the operator more flexibility in dealing with animals that have compromised lung function. But during anesthesia most animal patients are lung healthy and the addition of this control can add to the complexity of setting the ventilator. For example, consider that you have a hypocapnic animal patient. To bring that animal patients’ CO2 level to a more normal range you must decrease the overall gas exchange, the Minute Volume (MV). Without this option, only one knob controls the MV, that is the Volume control. With this option, however, increasing the I:E ratio would also cause a reduction in MV.
Hallowell Model 2024IE Pro
The Model 2024IE Pro is a 300-1600ml microprocessor-controlled unit providing the latest in electronic updates. Developed as a compliment to your already owned 2002 PRO, all parts, accessories, and consumables for the Model 2024PRO and Model 2002PRO are completely interchangeable.
All Hallowell PRO series ventilator purchases now include a Ventilator Accessory Kit.
PART # | DESCRIPTION |
000A6761 | 2024IEPRO Anesthesia Ventilator (0-300mL bellows, housing & adapter) |
000A6762 | 2024IEPRO Anesthesia Ventilator (300-1600mL bellows & housing) |
000A6763 | 2024IEPRO Anesthesia Ventilator (1600-3000mL bellows & housing) |
000A6764 | 2024IEPRO Anesthesia Ventilator (300-1600mL bellows & housing with 0-300mL attachment) |
If you have any questions, need more information on a product or don’t see a product you are looking for, please call us, email us or you can request pricing and we will be happy to assist you in any way we can.